pH Range pH: -2 ~ 16.00 / Accuracy ± 0.01 + 1 digit
Temperature range: 0 ~ 90°C / Accuracy ± 0.2°C
ATC included. Auto shut off after 10 minutes. Batteries 2 x CR232
PREPARATION
Remove the protective cap and unscrew the inner cap from the meter and rinse the electrode with clean water, then wipe dry.
Press the green power CAL button to turn the meter on. Always replace protective caps after usage. Ensure inner cap has pH 4 inside for electrode storage.
SINGLE POINT CALIBRATION
Check that the PH icon shows in the LCD display.
Dip the electrode into buffer 7.0 and stir gently, then wait for the display to stabilise.
Press and hold the green power CAL button to enter calibration mode and hold until the display shows CAL and then flashes 7.00. When the display stops flashing and indicates SA then END calibration is finished and the meter returns to measurement mode.
TWO OR THREE POINT CALIBRATION
After rinsing the electrode in clean water wipe dry.
Then dip the electrode into bu er 4.0 and stir gently, then wait for the display to stabilise.
Press and hold the green power CAL button to enter calibration mode and hold until the display shows CAL and then flashes 4.00. When the display stops flashing and indicates % then SA then END calibration is finished and the meter returns to measurement mode.
After slope calibration pH 4 or pH 10, the display will indicate percentage of slope to show status of the electrode.
If the % is below 70% or above 130%, the electrode must be replaced. A slope of 100% is ideal.
Notes: If the icon SA does not appear, then the calibration has failed.
When doing a 2 or 3 point calibration, use pH 7 first, followed by pH 4 and then pH 10.
MEASUREMENT
After calibration, rinse the electrode with clean water and wipe it dry.
Dip the electrode into the sample and stir gently until a stable reading is obtained. After reading, rinse in clean water and replace both caps.
Features & Benefits
- Easy to use, compact and convenient design
- Testing Chlorine LR
- Measuring Total Hardness
- Testing turbidity
- Testing conductivity/TDS
- Microbial monitoring using TTC/E.coli and TTC/Pseudomonas dip slides
- Testing pH
- Comes with Discovery Incubator